Deployment NameGliderProjectDeployedRecoveredDaysDistancePayload BayOSNotes
ru30-20191015T1824ru30OTIC-pH/Hurricane
Gliders
2019-10-15
18:24Z
2019-11-07
00:11Z
22.2471.4 km94178.20

All Notes (17)

Created by: dkaragon @ 2021-06-14 19:55:46Informational

li ion : pH

Cell Serial Volts Charge_AH Capacity_AH Percent_Charge Percent_Mean_Capacity
1 501 14.309 3.664 10.084 36 100
2 530 14.311 3.649 10.011 36 99
3 545 14.31 3.639 10.013 36 99
4 553 14.307 3.573 9.95 36 99
5 556 14.312 3.624 9.986 36 99
6 574 14.302 3.662 10.046 36 100
7 581 14.305 3.679 10.12 36 100
8 585 14.304 3.518 9.879 36 98
9 599 14.303 3.618 9.983 36 99
10 602 14.304 3.71 10.231 36 102
11 605 14.304 3.776 10.285 37 102
12 611 14.303 3.597 10.009 36 99
13 614 14.301 3.508 9.937 35 99
14 616 14.308 3.569 9.994 36 99
15 625 14.305 3.644 10.067 36 100
16 628 14.304 3.574 10.008 36 99
17 637 14.302 3.596 9.992 36 99
18 640 14.307 3.75 10.235 37 102
19 641 14.304 3.715 10.053 37 100
20 642 14.306 3.649 10.07 36 100
21 643 14.304 3.798 10.274 37 102
22 654 14.306 3.769 10.194 37 101
23 655 14.305 3.821 10.304 37 102
24 655 14.305 3.821 10.304 37 102

Created by: dkaragon @ 2021-02-10 16:49:49Alert

FLBBCD BS channel high on checkout for 2021 OA 1
(perhaps this deployment or earlier?)

Created by: dkaragon @ 2019-12-10 16:56:10Informational

Lithium post recovery analysis:

136 AH used by glider
Cells were about 3.65 AH / 10.08 AH -- (if we account for SD rate @ .3%/day then we had about 4 AH/cell at recovery (this note about 30 days past recovery)
4 AH remaining thus about 6/10 burned for mission --> 6 * 23 = 138 AH roughly matching glider. If we didn't account for SD rate it would be about 146 AH used

Thus glider reads 1 to 7% lower than batteries. Fairly good agreement as it might be more towards the 1% than 7%.

With 10% remaining on cells as a buffer, about 1 AH we had a remaining 2.65 AH on these cells. 2.65 * 23 = 60 AH. 136 + 60 = 195 total AH.

Created by: dkaragon @ 2019-11-06 19:30:14Informational
Modified by: dkaragon @ 2019-11-07 02:05:25

Recovered by Nicole, Chip, Liza, Sal out of Morgan Marina aboard RV Rutgers. CTD cast + rosette performed at recovery. CTD 1645, rosette CTD 1383

Created by: dkaragon @ 2019-11-06 18:19:46Informational

First rosette attempt:
MissionSTARTDate: 06 Nov 2019 18:17:04 Z
Mission Name: 200_N.MI
Mission Number: ru30-2019-309-2-0 (0356.0000)

Created by: dkaragon @ 2019-11-06 18:19:38Informational

Found a plume strange fail to surface segment 11/5 pm / 11/6 am

Created by: dkaragon @ 2019-10-28 18:22:28Informational

Still no response from optode:

# proglet = oxy4
# uart = 1 # U4Soem Pins T-2,R-3
# bit = 2 # power, 8.8v
# start_snsr = c_oxy4_on(sec)

SciDos>u4stalk 1 9600 2
Drop Carrier Detect for 3 seconds (i.e. unpower freewave)
--or--Type Ctrl-C and hit NO keys for 1 secs.
----------------------------------------------------------------------

Created by: dkaragon @ 2019-10-25 15:08:26Action Taken
Modified by: dkaragon @ 2019-10-25 17:02:05

Likely optode errors, u4stalk failed at surface.

Should try again in a few days. Right now disabled in proglets.dat

Sending back on to shelf before south. Strong currents and other reasons.

Created by: waite @ 2019-10-24 13:20:49Operations

pH line to shelf break complete. Deployment being shared between pH and Hurricane projects. Switching over to hurricane triangle now.

Created by: dkaragon @ 2019-10-23 12:07:32Informational

22 deg upcasts to help slow ascents

Created by: dkaragon @ 2019-10-16 23:09:37Action Needed

f_battpos_db_frac_fz .75 from .5

.5 guess is no good, noticed some pitch oddities

Should look to see why its .5, or if that is normal?

Created by: dkaragon @ 2019-10-16 22:43:47Informational

loaded mission ru30.mi

Optode, CTD, FLBBCD all off at surface now. pH on at surface

Back to normal swimming.

Created by: waite @ 2019-10-16 14:38:52Informational

# ru30
# 2019/10/16 waite@marine pH sensor cal values


####################################
# pH SBE41N SN C18
# 2019/8/01 waite@marine verified
sensor: sci_sbe41n_ph_is_installed(bool) 1 # in, t--> installed on science
sensor: c_sbe41n_ph_num_fields_to_send(nodim) 5 # in, number of columns to send on each
sensor: u_sbe41n_ph_is_calibrated(bool) 1 # in, false, assume not calibrated
sensor: u_sbe41n_ph_f0(Mnodim) -2885.6 # f0 -2.8856E-03
sensor: u_sbe41n_ph_f1(Mnodim) 15.063 # f1 1.5063E-05
sensor: u_sbe41n_ph_f2(Mnodim) -0.019683 # f2 -1.9683E-08
sensor: u_sbe41n_ph_f3(Mnodim) 0.000014037 # f3 1.4037-11
sensor: u_sbe41n_ph_f4(Mnodim) -0.0000000050132 # f4 -5.0132E-15
sensor: u_sbe41n_ph_k0(nodim) -1.393019 # k0 -1.393019
sensor: u_sbe41n_ph_k2(Mnodim) -1059.0 # k2 -1.059E-03
####################################

Created by: dkaragon @ 2019-10-15 19:57:46Informational

Thruster had an error on deck, < 20% might be too little? Seemed to work on test missions.

Rosette cast:
the8x3_filename: 03460000
full_filename: ru30-2019-287-2-0
NBD downloaded from that segment

0345 was shallow mission, 47, 48 were thruster test missions

4 hour mission to 10 m below to cross shipping lanes.

Created by: dkaragon @ 2019-10-15 19:56:24Informational
Modified by: dkaragon @ 2019-10-16 13:58:34

Deployed by Nicole Waite, Grace Saba, Chip Haldeman, & OMDA class out of Morgan Marina aboard RV RUTGERS.

CTD used was 1383 pumped SB19 on SB55 Rosette Nisky Sixx (segment 03460000) for comparison and pH water samples.

CTD 1645 pumped was used with first shallow mission (segment 03450000) for CTD comparion and profile to pick pH water sample depths.

Created by: kerfoot @ 2019-10-15 19:55:56Operations

Deployed by Rutgers University OMDA class cruise by Chip Haldeman, Grace Saba, Nicole Waite and lots of students aboard the R/V Rutgers.

Created by: kerfoot @ 2019-10-15 19:54:27Informational

OS 8.2
CTD S/N 9417 (cal'd 2019-08-07)
FLBBCD S/N 4662 (cal'd 2017-02-23)
pH C18 (cal'd 2019-08-01)
oxy4 S/N 242 (cal'd 2019-09-26)