Fleet Configurations
Deployments
Gliders
Projects
Payload Bays
Instruments
Users
API
Log In
Deployment ru34-20220330T1435
Pilot Notes
this note was for previous deployment, accidentally put it here... but something to keep an eye out for... "Optode foil was shifted at check-in. Could see the light at the edge of the sensing window. This was causing high readings in 100% bath. After shifting foil back to cover the light, readings were better, 101%. Data from deployment may have been impacted by this shifting foil. Prior communications with YSI said that a shifted foil and being able to see the light can cause wonky readings."
Category
Informational
Alert
Action Needed
Action Taken
Operations
Update
Credentials Required
Username
Password