Created by: dkaragon @ 2025-04-23 16:29:02Action Taken

MRI Service 4/2025

Summary:
Your instrument is in good condition and does not require any major repairs, although it will need some small repairs listed in the page 2 of the RMA report. Also, no faults were found that could have caused the issue reported with the CSR-2100, as stated in page 8.

Regarding the sensors, 1x shear (M2372) and 1x FP07 (T1427) failed inspection and require replacement. We’ve included 1x FP07-1000 with guard which has more protect (see attached image) as a replacement for T1427.

Created by: dkaragon @ 2023-01-23 00:18:53Informational

Leak was from a forward pump screw near top (bellows screw)

Created by: dkaragon @ 2023-01-15 00:31:35Informational

Glider triggered a leak in the front section at the tail end of a climb. Drifted. Recovered by Nicole Waite aboard RV Gould.

Created by: lgarzio @ 2023-01-13 17:26:52Informational

OS 8.6 Beta
Seabird CTD sn 9395 cal 6/19/2022
Wetlabs FLBBCD-SLC sn 3320 cal 2/12/2020
Rockland Scientific Microrider sn 410 cal n/a

Created by: lgarzio @ 2023-01-13 17:25:35Operations

Deployed by Nicole Waite, Carlos Moffat, Frederike Benz, Michael Cappola aboard Laurence M Gould near Adelaide Island, Antarctica