Fleet Configurations
Deployments
Gliders
Projects
Payload Bays
Instruments
Users
API
Log In
Deployment ru29-20221005T1517
Pilot Notes
Once glider surfaced from initial leak, leak detect started to slowly climb. After 35 minutes, it made it’s way above 2 and then after sending the mbd (so maybe ~1 hour after surfacing) it was up to 2.39. Seemed to level off around 2.4, was high 2.4's throughout deployment thus far but that's more likely noise on the AD circuit. Minimum value after abort was triggered was 1.31. Triggering forward leak detect while on the climb at around 140 meters is interesting. My theory thus far is that once the glider got to the surface, it warmed up, evaporating some of the water and the dessicant absorbed the moisture. Testing driving now to gradually deeper depths. So far 20 and 50 meters dives have not caused another abort.
Category
Informational
Alert
Action Needed
Action Taken
Operations
Update
Credentials Required
Username
Password