        "attrs": {
            "calibration_date": "2022-02-23",
            "comment": "",
            "description": "A conductivity, temperature and pressure sensor designed for use on gliders and autonomous underwater vehicles (AUVs). It can be used to derive salinity, density and sound velocity and is optimised for turbulence measurements and passive acoustic monitoring due to its silent operation (non-pumped). The instrument is available as two configurations; the standard RBR Legato3 has a 2 Hz standard thermistor response; the RBR Legato3 Fast 16 instrument has a 16 Hz fast thermistor response. Additional sensors such as dissolved oxygen, fluorescence, turbidity, and PAR are all supported. The RBR Legato3 is designed to fit into a standard 2 x 7 inch sensor bay, with both wet- and dry-bay models. The RBR Legato3 is depth-rated to 1000m, with a pressure range from 500-1000 db and accuracy of +/- 0.05 percent (full scale). The conductivity accuracy is +/- 0.003 mS/cm. Temperature range is -5 degC to +42 degC (ITS-90) with an accuracy of +/- 0.002 degC (-5 to +35 degC) and +/- 0.004 degC (+35 to +42 degC).",
            "factory_calibrated": "2022-02-23",
            "long_name": "RBR Legato3 CTD",
            "make_model": "rbrlegato3",
            "maker": "RBR",
            "maker_vocabulary": "https://vocab.nerc.ac.uk/collection/L35/current/MAN0049/,https://vocab.nerc.ac.uk/collection/B75/current/ORG00299/",
            "model": "RBR Legato3 CTD",
            "model_vocabulary": "https://vocab.nerc.ac.uk/collection/L22/current/TOOL1745/",
            "owner": "Rutgers University",
            "references": "",
            "serial_number": "203789",
            "type": "CTD,salinity sensor,water temperature sensor,water pressure sensor",
            "type_vocabulary": "http://vocab.nerc.ac.uk/collection/L05/current/130/,http://vocab.nerc.ac.uk/collection/L05/current/350/,http://vocab.nerc.ac.uk/collection/L05/current/134/,http://vocab.nerc.ac.uk/collection/L05/current/WPS/",
            "user_calibrated": ""
        },
        "cfg_file": "/home/glideradm/doco/glideradm/configs/instruments/203789-rbrctd.yml",
        "nc_var_name": "instrument_rbrctd",
        "type": "i4"
    },
    {
        "attrs": {
            "calibration_date": "2021-08-31",
            "comment": "Chlorophyll a,beta700nm,CDOM",
            "description": "A variant of the ECO Puck Triplet (http://vocab.nerc.ac.uk/collection/L22/current/TOOL0673/). It is a three-optical-sensor, user-defined instrument that carries a chlorophyll-a fluorometer, FDOM fluorometer and scattering meter. ECO Pucks are real-time only sensors as they are integrated onto the OEM platform that provides power and data handling. The SLC designation signifies this is a third generation model that is specific for integration into Slocum gliders with a dry science bay. The fluorometers and scattering meter are single wavelength sensors. The model is fitted with an added copper faceplate for bio-fouling mitigation and is depth-rated to 1000 m.",
            "factory_calibrated": "2021-08-31",
            "long_name": "ECO Puck Triplet FLBBCD-SLC",
            "maker": "WET Labs,Sea-Bird Scientific",
            "maker_vocabulary": "http://vocab.nerc.ac.uk/collection/L35/current/MAN0026/,http://vocab.nerc.ac.uk/collection/B75/current/ORG00226/",
            "model": "WETLabs ECO Puck Triplet FLBBCD-SLC scattering fluorescence sensor",
            "model_vocabulary": "https://vocab.nerc.ac.uk/collection/L22/current/TOOL1312/",
            "owner": "Rutgers University",
            "serial_number": "4662",
            "type": "sensor_model,optical backscatter sensor,fluorometer",
            "type_vocabulary": "http://vocab.nerc.ac.uk/collection/L19/current/SDNKG17/,http://vocab.nerc.ac.uk/collection/L05/current/123/,http://vocab.nerc.ac.uk/collection/L05/current/113/",
            "user_calibrated": ""
        },
        "cfg_file": "/home/glideradm/doco/glideradm/configs/instruments/4662-flbbcdslc.yml",
        "nc_var_name": "instrument_flbbcdslc",
        "type": "i4"
    },
    {
        "attrs": {
            "calibration_date": "2023-06-19",
            "comment": "Model 4831. In-house calibration 2024-10-15.",
            "description": "A dissolved oxygen sensor which provides analogue (0-5V) and digital output (RS-232) to third party data loggers, gliders and floats. Measurement based on the ability of selected substances to act as dynamic fluorescence quenchers. The fluorescent indicator is a special platinum porphyrin complex embedded in a gas permeable foil that is exposed to the surrounding water. In this standard model, a black optical isolation coating protects the complex from sunlight and fluorescent particles in the water. This sensing foil is attached to a window providing optical access for the measuring system from inside a watertight housing. The foil is excited by modulated blue light, and the phase of a returned red light is measured. For improved stability the 4831 optode also performs a reference phase reading by use of a red LED that does not produce fluorescence in the foil. This model is fitted with a standard sensing foil. By linearizing and temperature compensating, with an incorporated temperature sensor, the absolute O2 concentration can be determined. Accuracy +/- 1.5% or 2uM; precision +/- 0.2 uM.",
            "factory_calibrated": "2023-06-19",
            "long_name": "AA Optode 4831",
            "maker": "Aanderaa Data Instruments AS",
            "maker_vocabulary": "http://vocab.nerc.ac.uk/collection/B75/current/ORG00228/,http://vocab.nerc.ac.uk/collection/L35/current/MAN0007/",
            "model": "Aanderaa Oxygen Optode 4831",
            "model_vocabulary": "http://vocab.nerc.ac.uk/collection/L22/current/TOOL1239/",
            "owner": "Rutgers University",
            "references": "",
            "serial_number": "644",
            "type": "sensor_model,dissolved gas sensors",
            "type_vocabulary": "http://vocab.nerc.ac.uk/collection/L19/current/SDNKG17/,http://vocab.nerc.ac.uk/collection/L05/current/351/",
            "user_calibrated": "2024-10-15"
        },
        "cfg_file": "/home/glideradm/doco/glideradm/configs/instruments/644-optode.yml",
        "nc_var_name": "instrument_optode",
        "type": "i4"
    }
]