Pilot Notes
Optode foil was shifted at check-in. Could see the light at the edge of the sensing window. This was causing high readings in 100% bath. After shifting foil back to cover the light, readings were better, 101%. Data from deployment may have been impacted by this shifting foil. Prior communications with YSI said that a shifted foil and being able to see the light can cause wonky readings.